Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
10-1
Chapter 10
Model 5014
i
to 5030
i
Upgrade
Procedures
This chapter describes the 5014
i
to 5030
i
upgrade procedures. For details,
see the following topics:
●
“
Instrument Preparation
” on page
10-1
●
“
SHARP Connection Bracket Assembly Installation
” on page
10-2
●
“
Radius Tube Adapter Installation
” on page
10-5
●
“
SHARP External Valve Installation
” on page
10-7
●
“
SHARP Optics Assembly Installation
” on page
10-9
●
“
Upgrade of the Heater Tube Assembly
” on page
10-10
●
“
Model Label Change
” on page
10-11
●
“
Power On Instrument
” on page
10-11
Use the following procedure to prepare the instrument.
1.
If upgrading at an installed site, take the appropriate measures to secure
the sample downtube and remove the instrument heater assembly and
set aside.
2.
From Instrument Panel, select Main Menu, select Instrument Controls
and then Service Mode. Toggle to “Service Mode On”.
3.
From Instrument Panel, select Filter Tape, and then Manual, and select
the BENCH to open.
4.
From Instrument Panel, select Diagnostics, Instrument Configuration,
and change the DETECTOR configuration from BETA to SHARP.
5.
Turn instrument power off with the measurement head in the open
position.
Instrument
Preparation
Содержание 5030i SHARP
Страница 4: ......
Страница 10: ...About This Manual Where to Get Help vi Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 168: ......
Страница 218: ......
Страница 294: ......
Страница 306: ......
Страница 402: ......
Страница 403: ......