Troubleshooting
Troubleshooting Guides
Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
6-3
Malfunction
Possible Cause
Action
LCD cable loose
Check connection and cable
integrity.
LCD defective
Replace display.
The change of
parameters is not
possible
Keyboard not enabled
Enable keyboard through menu.
No mass value
ADC defect
Change the main circuit board.
Mains are broken
Check and repair.
No high voltage value
Check HV.
Detector defect
Change detector.
No automatic filter
change
Mechanical defect
Check filter transport motor and
mechanics. Make sure tape is
adhered to take-up reel and
intact.
If pump is not running
Check pump power supply and
wires.
If the filter change parameter
reaches the FC-value for the
second time, then, the monitor
stops
Adjust filter change parameters.
LCD out of function or
dark
Wrong contrast
Adjust contrast.
LCD defect
Change LCD.
Connection cables from the
main board are broken
Replace display ribbon cable.
Air flow rate
disturbance or air flow
is too low
Air flow parameters wrong
Check air flow rate parameters.
No connection with plug (25)
“pump”
Put in the plug (pump regulation
input).
Pump do not work properly
Clean inside parts of pump
change.
Very high pollution or blocking
in the air passage
Clean air passage.
Check
P calibration valve are
in closed position
Close valves.
Neph Board Failures
External cable not connected
Check cable connection.
Internal cable not connected.
Check cable connection.
Low Source Current
Disconnected Source/Ref
cable
Check cable connection.
Содержание 5030i SHARP
Страница 4: ......
Страница 10: ...About This Manual Where to Get Help vi Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 168: ......
Страница 218: ......
Страница 294: ......
Страница 306: ......
Страница 402: ......
Страница 403: ......