C-Link Protocol Commands
Diagnostics
Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
B-25
Send:
diag
volt
mb
Receive:
diag volt mb 24.10 14.90 4.90 3.20 -3.20
diag volt mib
This command reports the diagnostic voltage measurements on the
measurement interface board. The sequence of voltages is: Positive 24,
positive 15, negative 15, positive 5, and positive 3.3. Each voltage value is
separated by a space.
Send:
diag
volt
mib
Receive:
diag volt mib 24.98 14.80 -14.90 4.96 3.20
diag volt neph
This command reports the current voltage readings of the nephelometer
interface board. The following example reports the nephelometer interface
board values.
Send:
diag
volt
neph
Receive:
diag volt mib 3.3 3.3 5.0 7.2 -5.0 -7
flow
This command reports the current sample flow. The following example
reports that the sample flow is 16.69 LPM.
Send:
sample
flow
Receive:
sample flow 16.69 LPM
flow pres
This command reports the current flow pressure for the detector, in
mmHg. The following example reports that the flow pressure is 24.1
mmHg.
Send:
flow
pres
Receive:
flow pres 24.1 mmHg
flow temp
This command reports the current flow temperature for the detector, in
degrees C. The following example reports that the flow temperature is 22.3
°C.
Send:
flow
temp
Receive:
flow temp 22.3 deg
io bd
This command reports the presence of the I/O expansion board (yes/no).
The following example reports the I/O expansion board is present (yes).
Содержание 5030i SHARP
Страница 4: ......
Страница 10: ...About This Manual Where to Get Help vi Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 168: ......
Страница 218: ......
Страница 294: ......
Страница 306: ......
Страница 402: ......
Страница 403: ......