Servicing
Fuse Replacement
7-14
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
Use the following procedure to replace the fuse.
Equipment Required:
Replacement fuses (refer to the “
Replacement Parts List
” in this
chapter).
1.
Turn the instrument OFF and unplug the power cord.
2.
Remove the fuse drawer, located on the AC power connector.
3.
If either fuse is blown, replace both fuses.
4.
Insert fuse drawer and re-connect power cord.
Use the following procedure to replace the fan and the fan filter (
Figure 7–
6
).
Equipment Required:
Fan
Fan filter
Philips screwdriver
Cable cutter (for tie-wrap removal)
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
▲
1.
Turn the instrument OFF, unplug the power cord, and remove the
cover.
2.
Remove the fan guard and filter from the fan by unsnapping it.
3.
If the fan is not being replaced, install the new filter, and snap it back
into place, and skip the remaining steps.
4.
Disconnect the fan power cable from the fan. Do not lose the four
nuts.
Fuse Replacement
Fan/Filter
Replacement
Содержание 5030i SHARP
Страница 4: ......
Страница 10: ...About This Manual Where to Get Help vi Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 168: ......
Страница 218: ......
Страница 294: ......
Страница 306: ......
Страница 402: ......
Страница 403: ......