Servicing
Analog Output Testing
7-22
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
1.
Connect a meter to the channel to be tested.
Figure 7–11
shows the
analog output pins and
Table 7–4
identifies the associated channels.
2.
From the Main Menu, choose Diagnostics >
Test Analog Outputs
.
The Test Analog Outputs screen appears.
3.
Press
to scroll to the desired channel corresponding to the rear
panel terminal pins where the meter is connected, and press
.
The Set Analog Outputs screen appears.
4.
Press
to set the output to zero.
The Output Set To line displays Zero.
5.
Check that the meter is displaying the zero value. If the meter reading
differs by more than one percent of the full-scale output, the analog
outputs should be adjusted. Refer to the “
Analog Output Calibration
”
procedure that follows.
6.
Press
to set the output to full-scale.
The Output Set To line displays Full-Scale.
7.
Check that the meter is displaying a full-scale value. If the meter
reading differs by more than one percent of the full-scale output, the
analog outputs should be adjusted. Refer to the “
Analog Output
Calibration
” procedure that follows.
8.
Press
to reset the analog outputs to normal.
Содержание 5030i SHARP
Страница 4: ......
Страница 10: ...About This Manual Where to Get Help vi Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 168: ......
Страница 218: ......
Страница 294: ......
Страница 306: ......
Страница 402: ......
Страница 403: ......