Servicing
SHARP RH/Temp Sensor Replacement
7-46
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
Use the following procedure to replace the SHARP RH/Temp sensor
(
Figure 7–26
).
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
▲
1.
Turn the instrument OFF, unplug the power cord, and remove the
nephelometer from the base instrument.
2.
Remove the two side cover screws, and remove cover.
3.
Disconnect the RH/Temp connector from the SHARP MIB board
“TEMP/RH”.
4.
Remove the two socket-head screws from the RH/Temp board
assembly from the upper optic chamber endplate.
5.
Remove the RH/Temp gasket from the old RH/Temp board assembly
and add it to the new RH/Temp board assembly.
6.
Slide in the new RH/Temp assembly by following the previous steps.
7.
Calibrate the RH/Temp sensor. Refer to the “
Calibration
” chapter in
this manual.
SHARP RH/Temp
Sensor
Replacement