Calibration
Calibration Procedure
Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
4-13
The Nephelometer Calibration submenu is used to view and set the
nephelometer calibration. The nephelometer calibration screen is visible
only when the instrument is in service mode. For more information on the
service mode, see “
Service Mode
” earlier in the chapter.
Note
This adjustment should only be performed by an instrument service
technician.
▲
●
In the Main Menu, choose Service >
Nephelometer Calibration
.
NEPHELOMETER CALIBRATION:
>NEPH RH 0.0 %
NEPH TEMP 0.0
o
C
NEPH SRC LEVEL 0 %
RANGE AVG DIAGS ALARM
The Nephelometer RH screen allows the user to view and calibrate the
nephelometer relative humidity.
Note
Calibration of the nephelometer RH must be done with the heater in
the “OFF” position.
▲
●
In the Main Menu, choose Service > Neph Calibration >
Neph RH
.
CALIBRATE NEPH RH:
NEPH RH: 0.0 %
OFFSET: 0.0 %
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Nephelometer Temperature screen allows the user to view and
calibrate the nephelometer temperature sensor calibration.
Note
Calibration of the nephelometer temperature must be done with the
heater in the “OFF” position.
▲
Nephelometer
Calibration
Nephelometer RH
Nephelometer
Temperature
Содержание 5030i SHARP
Страница 4: ......
Страница 10: ...About This Manual Where to Get Help vi Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 168: ......
Страница 218: ......
Страница 294: ......
Страница 306: ......
Страница 402: ......
Страница 403: ......