Operation
Service Menu
Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
3-93
THERMAL MASS COEFFICIENT:
CURRENTLY: 0.000000
SET TO: 0023.00
0
?
MOVE CURSOR
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
The Vacuum Mass coefficient screen allows the user to view and set the
vacuum coefficient using manual calibration mode. (Factory set.)
●
In the Main Menu, choose Service > Mass Calibration >
Vacuum
Coefficient
.
VACUUM MASS COEFFICIENT:
CURRENTLY: 0.000000
SET TO: 0002.00
0
?
MOVE CURSOR
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
The Barometer Mass Coefficient screen allows the user to view and set the
barometer coefficient using manual calibration mode. (Factory set.)
●
In the Main Menu, choose Service > Mass Calibration >
Baro
Coefficient
.
BARO. MASS COEFFICIENT:
CURRENTLY: 0.000000
SET TO: 0.00054
0
?
MOVE CURSOR
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
The Mass Coefficient submenu allows the user to view and set the mass
coefficient using manual or auto mode. (Factory set.)
●
In the Main Menu, choose Service > Mass Calibration >
Mass
Coefficient
.
Vacuum Mass Coefficient
Barometer Mass
Coefficient
Mass Coefficient
Содержание 5030i SHARP
Страница 4: ......
Страница 10: ...About This Manual Where to Get Help vi Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 168: ......
Страница 218: ......
Страница 294: ......
Страница 306: ......
Страница 402: ......
Страница 403: ......