Model 5014i to 5030i Upgrade Procedures
SHARP Connection Bracket Assembly Installation
10-2
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
6.
Unplug instrument from AC power source.
INSTRUMENT CONFIGURATION:
>I/O EXPANSION BOARD YES
DETECTOR BETA
DETECTOR B NONE
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
Use the following procedure for SHARP connection bracket assembly
installation.
1.
Locate the Connection Bracket (
Figure 10–1
) and remove two screws
and set aside.
4-pin SHARP
Connector
2-pin connector
not used
Figure 10–1.
SHARP Optics Connection Bracket
2.
Remove both sides of the Cover. Remove the tape pick up spool
(
Figure 10–2
) by removing the thumb screw, washer, filter tape,
spacers, o-rings and rear filter support guide plate.
SHARP Connection
Bracket Assembly
Installation