Model 5014i to 5030i Upgrade Procedures
Radius Tube Adapter Installation
10-6
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
1 = UNFASTEN 2 SCREWS & REMOVE BRACKET
2 = SLIDE BACK BRACKET DOWNWARDS
3 = WITH TWISTING MOTION REMOVE THE EXISTING THREADED FACE
SEAL AND USE SAME MOTION TO INSTALL RADIUS TUBE ADAPTER
1
3
2
Figure 10–5.
Radius Tube Adapter Install
6.
Reassemble the two covers of the instrument in reverse order and check
alignment of radius tube adapter through the hole in the cover as
shown in
Figure 10–6
. Be sure to remove the old dust cover cap and
the large plastic hole cover.
Figure 10–6.
Radius Tube Adapter Alignment Check
Содержание 5030i SHARP
Страница 4: ......
Страница 10: ...About This Manual Where to Get Help vi Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 168: ......
Страница 218: ......
Страница 294: ......
Страница 306: ......
Страница 402: ......
Страница 403: ......