Calibration
Calibration Procedure
4-10
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
3.
Cut/break the filter tape and remove from the bench. Then insert the
Foil Holder from the calibration kit and press
while maintaining
a slight sideward pressure on the filter holder.
MASS COEF: 7000.0
FOIL VALUE: 132
8
3
μ
g
REMOVE TAPE, INSERT FOIL
HOLDER,
TO CLOSE BENCH
RANGE AVG DIAGS ALARM
4.
Insert the Null/Zero foil with the label side up and press
to
proceed with the foil calibration, or press
to stop the foil.
●
●
●
●
The auto zero procedure of mass will begin and last approximately 270
seconds. When the zero is complete, the beta attenuation measurement
will remain closed and an average Beta count rate is displayed.
MASS COEF: 7000.0
FOIL VALUE: 132
8
3
μ
g
Beta Avg 12022 1/sec
REMOVE NULL FOIL
TO CONTINUE
RANGE AVG DIAGS ALARM
5.
Remove the zero/null foil and insert with the Span foil (e.g., 1328 ug)
and press
to begin the span calibration, or press
to stop
the foil.
MASS COEF: 7000.0
FOIL VALUE: 132
8
3
µ
g
INSERT NULL FOIL
TO START ZERO CAL
RANGE AVG DIAGS ALARM
MASS COEF: 7000.0
FOIL VALUE: 132
8
3
µ
g
Beta Avg 12016 1/sec
PLEASE WAIT 265 sec
RANGE AVG DIAGS ALARM
Содержание 5030i SHARP
Страница 4: ......
Страница 10: ...About This Manual Where to Get Help vi Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 168: ......
Страница 218: ......
Страница 294: ......
Страница 306: ......
Страница 402: ......
Страница 403: ......