Servicing
Internal SHARP Cable Assembly Replacement
Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
7-51
Use the following procedure to replace the case cable (
Figure 7–30
).
Equipment Required:
Internal SHARP Cable Assembly
Philips screwdriver
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
▲
1.
Turn the instrument OFF and unplug the power cord.
2.
If the instrument is mounted in a rack, remove it from the rack.
3.
Refer to “
Removing the Measurement Case Assembly and Lowering
the Partition Panel
” in this chapter to lower the partition panel, then
proceed to the next step below.
4.
Disconnect the electrical connector labeled “SHARP 1”.
5.
Unfasten the four screws.
6.
Pull connect up and out and install electrical connector by following
the previous steps in reverse.
Internal SHARP
Cable Assembly
Replacement