Servicing
Ambient RH/Temperature Replacement
Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
7-39
1.
Turn the instrument OFF, unplug the power cord, and remove the
cover.
2.
Loosen the two screws of the radius mounting block using a 9/16-inch
hex driver.
3.
Remove radius tube SHARP adapter.
4.
Loosen the two mounting screws on the measurement head assembly.
Remove tube by slightly turning side-to-side, while pulling.
5.
Verify the sealing o-ring is in position before sliding the new radius
tube assembly into the measurement head.
6.
Install the new radius tube assembly by following the previous steps in
reverse.
Figure 7–21.
Replacing the Radius Tube
Use the following procedure to replace the ambient RH/temperature
assembly (
Figure 7–22
).
Equipment Required:
Ambient RH/temperature assembly
Adjustable wrench
Pliers
Ambient
RH/Temperature
Replacement
Unfasten Screws (2)
Pull Out Tube
Unfasten Screws (2)
Insulation
Radius Tube Adapter