Servicing
Ambient RH/Temperature Replacement
7-40
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
▲
1.
Turn the instrument OFF, unplug the power cord, and remove the
cover.
2.
Remove the cable from the bottom of the radiation shield assembly that
houses the ambient RH/temperature assembly.
3.
Remove the two mounting nuts on the mounting clamp of the
radiation shield assembly. Remove the radiation shield assembly.
4.
Turn over the radiation shield assembly and remove the three thumb
screws.
5.
Remove the three spacers and mounting bracket.
6.
Slide out the ambient RH/temperature assembly.
7.
Slide in the new ambient RH/temperature assembly by following the
previous steps in reverse.
8.
Calibrate the instrument. Refer to the “
Calibration
” chapter in this
manual.