Model 5014i to 5030i Upgrade Procedures
SHARP External Valve Installation
Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
10-7
Use the following procedure for SHARP external valve installation.
1.
Attach the SHARP External Valve Assembly (which includes the to the
Rear Panel using the captive hardware on bracket and attach the 3-pin
connector to the rear panel labeled valve (
Figure 10–7
).
Figure 10–7.
Rear Valve Assembly Installation
2.
Follow the following plumbing procedure in accordance with
Figure
10–8
below.
a.
Attach the 3/8” vacuum tubing between pump intake and the
vacuum port on rear panel and include the push-to-connect “T”
fitting with the 0.040” thru-hole on the T-side in this line.
b.
From the valve assembly that is now attached to the rear panel,
connect the threaded fitting to the pump exhaust.
SHARP External
Valve Installation
Port 1
Port 2
Содержание 5030i SHARP
Страница 4: ......
Страница 10: ...About This Manual Where to Get Help vi Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 168: ......
Страница 218: ......
Страница 294: ......
Страница 306: ......
Страница 402: ......
Страница 403: ......