Operation
Service Menu
Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
3-89
●
●
●
●
The Vacuum/Flow Calibration submenu is used to calibrate the
vacuum/flow offset, vacuum/flow span, or reset default values.
●
In the Main Menu, choose Service > Pres/Vacuum Calibration >
Vac/Flow
.
VAC PRES CALIBRATION:
>VAC/FLOW OFFSET
VAC PRES SPAN 1.0000
FLOW PRES SPAN 1.0000
SET DEFAULTS
RANGE AVG DIAGS ALARM
The Calibrate Vacuum/Flow Offset screen calibrates the vacuum/flow
sensor offset value.
●
In the Main Menu, choose Service > Pres/Vacuum Calibration >
Vac/Flow >
Vac/Flow Offset
.
CAL VACUUM/FLOW OFFSET:
VAC OFFSET: 0.0
FLOW OFFSET: 0.0
TO CALIBRATE ZERO BY
CHANGING FILTER
RANGE AVG DIAGS ALARM
The Calibrate Vacuum Pressure Span screen allows the user to view and set
the flow sensor calibration span point.
Note
The operator should use an independent manometer capable of
measuring approximately 100 mmHg of vacuum. Measure the vacuum
Vacuum/Flow Calibration
Calibrate Vacuum/Flow
Offset
Calibrate Vacuum
Pressure Span
RESTORE DEFAULT CAL:
RESTORE
RANGE AVG DIAGS ALARM
RESTORE DEFAULT CAL:
RESTORE
ARE YOU SURE YOU WANT TO?
x
PRESS
TO CONFIRM RESTORE
RANGE AVG DIAGS ALARM
Содержание 5030i SHARP
Страница 4: ......
Страница 10: ...About This Manual Where to Get Help vi Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 168: ......
Страница 218: ......
Страница 294: ......
Страница 306: ......
Страница 402: ......
Страница 403: ......