Operation
Service Menu
Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
3-85
The Service menu appears only when the instrument is in the service mode.
When the service mode is active, the service (wrench) icon is displayed on
the right side of the status bar. To put the instrument into the service
mode:
●
In the Main Menu, choose Instrument Controls >
Service Mode
.
Advanced diagnostic functions are included in the service mode.
Meaningful data should not be collected when the instrument is in the
service mode.
●
In the Main Menu, choose
Service
.
SERVICE:
>RH/TEMP CALIBRATION
PRES/VACUUM CALIBRATION
FLOW CALIBRATION
MASS CALIBRATION
DETECTOR CALIBRATION
RANGE AVG DIAGS ALARM
NEPH CALIBRATION
ANALOG OUT CALIBRATION
ANALOG INPUT CALIBRATION
DISPLAY PIXEL TEST
RESTORE USER DEFAULTS
RESTORE FACTORY DEFAULTS
The RH/Temperature Calibration submenu allows the user to view and set
the relative humidity and temperature sensor calibration. The
RH/temperature calibration submenu is visible only when the instrument is
in service mode. For more information on the service mode, see “
Service
Mode
” earlier in this chapter.
Note
This adjustment should only be performed by an instrument service
technician.
▲
●
In the Main Menu, choose Service >
RH/Temp Calibration
.
Service Menu
RH/Temperature
Calibration
Содержание 5030i SHARP
Страница 4: ......
Страница 10: ...About This Manual Where to Get Help vi Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 168: ......
Страница 218: ......
Страница 294: ......
Страница 306: ......
Страница 402: ......
Страница 403: ......