Servicing
Case Cable Replacement
7-52
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
Figure 7–30.
Replacing the Internal SHARP Cable Assembly
Use the following procedure to replace the case cable (
Figure 7–31
).
Equipment Required:
Case Cable
Philips screwdriver
Adjustable wrench
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
▲
1.
Turn the instrument OFF, unplug the power cord, remove the
electrical connector, loosen the two captive screws on the SHARP
optics base plate, and remove the SHARP assembly from the base unit.
2.
Remove the two SHARP cover screws and remove the cover.
3.
Remove the hex nut from the cable clamp.
Case Cable
Replacement
Screws (4)
“SHARP 1”
2 Pin (Not Used)