C-Link Protocol Commands
Alarms
Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
B-19
representation of the concentration alarm limits. Values must be in the
units that are currently set for use. The following example sets the instant
SHARP concentration alarm maximum value to 10000 μg/m
3
.
Send:
set alarm conc inst sharp max 10000
Receive:
set alarm conc inst sharp max 10000 ok
alarm filter tape status
This command reports the status of the filter tape as FAIL or OK. The
following example reports that the filter tape status is OK.
Send:
alarm filter tape status
Receive:
alarm filter tape status ok
alarm flow min
alarm flow max
These commands report the current flow alarm minimum and maximum
value settings. The following example reports that the flow alarm minimum
value is 16.0 LPM.
Send:
alarm flow min
Receive:
alarm flow min 16.00 LPM
set alarm flow min
value
set alarm flow max
value
These commands set the flow alarm minimum and maximum values to
value
, where
value
is a floating-point number representing flow alarm limits
in liters per minute/liters per hour. The following example sets the flow
alarm maximum value to 17.34 LPM.
Send:
set alarm flow max 17.34
Receive:
set alarm flow max 17.34 ok
alarm flow pres min
alarm flow pres max
These commands report the current flow pressure alarm minimum and
maximum value settings. The following example reports that the flow
pressure alarm minimum value is -10.0 mmHg.
Send:
alarm flow pres min
Receive:
alarm flow pres min -10.0 mmHg
set alarm flow pres min
value
set alarm flow pres max
value
These commands set the flow pressure alarm minimum and maximum
values to
value
, where
value
is a floating-point number representing flow
pressure alarm limits in millimeters of mercury. The following example sets
the flow pressure alarm maximum value to 50.0 mmHg.
Содержание 5030i SHARP
Страница 4: ......
Страница 10: ...About This Manual Where to Get Help vi Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 168: ......
Страница 218: ......
Страница 294: ......
Страница 306: ......
Страница 402: ......
Страница 403: ......