Operation
Alarms Menu
Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
3-71
the minimum or maximum limit, an alarm is activated and the alarm (bell)
icon appears in the status bar on the Run screen and in the Main Menu.
●
In the Main Menu, choose Alarms > Neph Alarms >
LED Current
.
LED SOURCE CURRENT:
ACTUAL 65.0 mA
>MIN 50.0 mA
MAX 70.0 mA
RANGE AVG DIAGS ALARM
The Minimum LED Source Current alarm limit screen is used to change
the minimum LED source current alarm limit. The minimum and
maximum LED Source Current screens function the same way.
●
In the Main Menu, choose Alarms > Neph Alarms > LED Current >
Min
or
Max
.
LED SOURCE CURRENT:
ACTUAL MIN: 50.0 mA
SET MIN TO: 50.1 mA?
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Reference Detector Voltage screen displays the current nephelometer
reference detector voltage value and the minimum and maximum alarm
limits. Acceptable alarm limits range 350 to 2000 mV. If the reference
detector voltage reading goes beyond the minimum or maximum limit, an
alarm is activated and the alarm (bell) icon appears in the status bar on the
Run screen and in the Main Menu. There is no “set” mode for this value.
●
In the Main Menu, choose Alarms > Neph Alarms >
Ref Det Volt
.
REF DETECTOR VOLTAGE:
ACTUAL 780 mV
MIN 350 mV
MAX 2000 mV
RANGE AVG DIAGS ALARM
Min and Max LED Source
Current Limits
Reference Detector
Voltage
Содержание 5030i SHARP
Страница 4: ......
Страница 10: ...About This Manual Where to Get Help vi Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 168: ......
Страница 218: ......
Страница 294: ......
Страница 306: ......
Страница 402: ......
Страница 403: ......