E-1
Appendix E
ESM Protocol Commands
This appendix provides a description of the ESM Protocol commands
(from the prior FH62 platform) that can be used to remotely control a
Model 5030
i
instrument using a host device such as a PC or a datalogger.
ESM protocol may be used over RS-232, RS-485 or over Ethernet. ESM
functions can be accessed over Ethernet using TCP port 9884.
Up to three simultaneous connections per protocol may be made over
Ethernet.
Table E–1
through
Table E–3
lists the ESM commands supported for the
Model 5030
i
.
Table E–1.
Read Commands for 5030
i
Command
Description
C3
PM conc always always μg/m³
C2
Neph conc μg/m³
C
SHARP conc μg/m³
H3
Avg PM conc always μg/m³
H2
Avg Neph conc μg/m³
HT
Avg SHARP conc μg/m³
JB
temperature sampling head–ambient temp
JD
temperature inside of the flow meter orifice–flow temp
JF
different pressure of the air flow measuring module in Pa
JG
low pressure suction chamber
JH
barometer in hPa
JI
air flow sampling head–volumetric flow
JJ
norm air flow (273K, 1013hPa)
JR
relative humidity value (%RH)
JS
sample relative humidity value (%RH)
m1
non filtered mass in μg or mg based on units selected
o
reports “offline” in service mode, “online” if NOT in service mode
ESM Commands
Supported
Содержание 5030i SHARP
Страница 4: ......
Страница 10: ...About This Manual Where to Get Help vi Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 168: ......
Страница 218: ......
Страница 294: ......
Страница 306: ......
Страница 402: ......
Страница 403: ......