Preventive Maintenance
Greasing the Cam
5-8
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
Figure 5–5.
Replacing the Filter Tape
Use the following procedure to grease the cam.
Equipment Required:
Dow Corning Molykote G-N
1.
Apply metal assembly paste around the parameter of the cam.
It is recommended to replace the external pump exhaust filter every six
months. On an annual basis, remove top plate of optics and vacuum out
optic chamber per the following procedure or send it for service and re-
calibration.
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component. For more
information about appropriate safety precautions, see the “
Servicing
”
chapter.
▲
Greasing the
Cam
External Pump
Exhaust Filter