Operation
Alarms Menu
3-76
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
BOARD TEMPERATURE:
ACTUAL MIN: 5.0
o
C
SET MIN TO: 0.0
o
C?
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Pressure/Vacuum Alarms submenu allows the user to view and set the
pressure alarm parameters.
●
In the Main Menu, choose Alarms >
Pressure/Vacuum Alarms
.
PRESSURE/VACUUM ALARMS:
>BARO PRES OK
VACUUM OK
FLOW OK
RANGE AVG DIAGS ALARM
The Barometric Pressure screen displays the current pressure reading and
sets the minimum and maximum alarm limits. Acceptable alarm limits
range from 400 to 800 mmHg. If the barometric pressure reading goes
beyond the minimum or maximum limit, an alarm is activated and the
alarm (bell) icon appears in the status bar on the Run screen and in the
Main Menu.
●
In the Main Menu, choose Alarms > Pressure/Vacuum Alarms >
Baro Pres
.
BAROMETRIC PRESSURE:
ACTUAL 764.0 mmHg
>MIN 400.0 mmHg
MAX 800.0 mmHg
RANGE AVG DIAGS ALARM
The Minimum Barometric Pressure alarm limit screen is used to change the
minimum barometric pressure alarm limit. The minimum and maximum
barometric pressure screens function the same way.
●
In the Main Menu, choose Alarms > Pressure/Vacuum Alarms > Baro
Pres >
Min
or
Max
.
Pressure/Vacuum
Alarms
Barometric Pressure
Min and Max Barometric
Pressure Limits
Содержание 5030i SHARP
Страница 4: ......
Страница 10: ...About This Manual Where to Get Help vi Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 168: ......
Страница 218: ......
Страница 294: ......
Страница 306: ......
Страница 402: ......
Страница 403: ......