Servicing
Proportional Valve Replacement
Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
7-33
4.
Remove board.
5.
Re-install the photo interrupt board by following the previous steps in
reverse.
Figure 7–16.
Replacing the Photo Interrupt Board
Use the following procedure to replace the proportional valve (
Figure 7–
17
).
Equipment Required:
Proportional valve assembly
Hex driver, 3/32-inch
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
▲
1.
Turn the instrument OFF, unplug the power cord, and remove the
cover.
2.
Refer to “
Removing the Measurement Case Assembly and Lowering
the Partition Panel
” in this chapter to lower the partition panel, then
proceed to the next step below.
3.
Disconnect connector from the measurement interface board.
Proportional Valve
Replacement
Unhook Cable
Unfasten screws (2)