Operation
Service Menu
3-96
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
DETECTOR AUTO CAL:
HIGH VOLT: 1500 V
BETA CNT : 14649 1/sec
PRESS
TO START AUTO CAL
RANGE AVG DIAGS ALARM
The Mass Coefficient submenu allows the user to perform a mass foil
calibration using a zero (null) and span foil set.
●
In the Main Menu, choose Service > Detector Calibration >
Manual
.
DETECTOR MANUAL CAL:
>BETA/REF/ALPHA COUNTS
HIGH VOLTAGE
BETA REF THRESHOLD
ALPHA THRESHOLD
RANGE AVG DIAGS ALARM
The Beta/Ref/Alpha Counts screen allows the user to view the real-time
alpha and beta counts.
●
In the Main Menu, choose Service > Detector Calibration > Manual
>
Beta/Ref/Alpha Counts
.
COUNTS:
alpha 0 1/sec
beta 10212 1/sec
beta corr 10145 1/sec
beta ref 5200 1/sec
RANGE AVG DIAGS ALARM
The High Voltage screen allows the user to view and set the high voltage
value.
●
In the Main Menu, choose Service > Detector Calibration > Manual
>
High Voltage
.
Manual Detector
Calibration
Beta/Ref/Alpha Counts
High Voltage