Operation
Alarms Menu
Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
3-79
FLOW ALARMS:
>FLOW OK
RANGE AVG DIAGS ALARM
The Flow screen displays the current flow and sets the minimum and
maximum alarm limits. Acceptable alarm limits range from 16.00 to 17.34
LPM. If the flow reading goes beyond the minimum or maximum limit, an
alarm is activated and the alarm (bell) icon appears in the status bar on the
Run screen and in the Main Menu.
●
In the Main Menu, choose Alarms > Flow Alarms >
Flow
.
FLOW:
ACTUAL 0.00 LPM
>MIN 16.17 LPM
MAX 17.17 LPM
RANGE AVG DIAGS ALARM
The Minimum Flow alarm limit screen is used to change the minimum
flow alarm limit. The minimum and maximum flow screens function the
same way.
●
In the Main Menu, choose Alarms > Flow Alarms > Flow >
Min
or
Max
.
FLOW:
ACTUAL MIN: 16.00 LPM
SET MIN TO: 16.17 LPM?
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Concentration Alarms submenu allows the user to view and set the 24-
hour average concentration and the instant concentration for PM, SHARP,
and nephelometer alarm parameters.
●
In the Main Menu, choose Alarms >
Conc Alarms
.
Flow
Min and Max Flow Limits
Concentration Alarms
Содержание 5030i SHARP
Страница 4: ......
Страница 10: ...About This Manual Where to Get Help vi Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 168: ......
Страница 218: ......
Страница 294: ......
Страница 306: ......
Страница 402: ......
Страница 403: ......