Function
111
[MENU] - [MAINTENANCE]
Performs the following maintenance/adjustment operation.
REAG.
Allows you to check the reagent volume and replace the reagent.
Item
Description
Reference
Allows you to check the reagent volume and replace the
reagent.
Page 111
Specifies the date of periodic inspection.
Page 115
Specifies the external analog output adjustment.
Page 116
Specifies the step action and allows you to execute it.
Page 117
Specifies and updates the separate action.
Page 119
Item
Setting range
Reagent capacity
Description
REAG VOLUME
CHECK
Reag. A
1000 mL
Allows you to check the remaining amount of
reagent.
Reag. B
250 mL
Reag. C
250 mL
Reag. D
250 mL
Reag. E
250 mL
Reag. F
250 mL
SPAN water
1000 mL
CHANGE REAG.
Total
-
Used when replacing all the reagent bottles at the
same time.
REAG VOLUME
INPUT
Reag. A
0 mL to 1000 mL
Allows you to specify the remaining amount of
reagent for each reagent when reagents are
replaced individually.
Reag. B
0 mL to 250 mL
Reag. C
0 mL to 250 mL
Reag. D
0 mL to 250 mL
Reag. E
0 mL to 250 mL
Reag. F
0 mL to 250 mL
SPAN water
0 mL to 1000 mL