Material
226
Specifications
General
specifications
Equipment name Automatic total nitrogen/phosphorus monitoring system
Model
TPNA-500
Dimensions
*1
460 (W) mm
385 (D) mm
1600 (H) mm
Mass
Approx. 87 kg
Power supply
100 V to 240 V AC
10
, 50 Hz/60 Hz
Power
consumption
100 V to 240 V AC: Approx. 250 VA
Performance
Measurement
range
*2
Total nitrogen: 0 mg N/L to 2/5/10/20/50/100/200/500/1000 mg N/L
Total phosphorus: 0 mg P/L to 0.5/1/2/5/10/20/50/100/250 mg P/L
Repeatability
Within
3
of the full scale
(Total nitrogen full scale: 50 mg N/L or less, total phosphorus full scale: 10 mg P/L or less)
Within
5
of the full scale
(Total nitrogen full scale: 100 mg N/L to 1000 mg N/L, total phosphorus full scale: 20
mg P/L to 250 mg P/L)
Both with standard solution
Measurement
principle
Total nitrogen: Alkaline potassium persulfate, ultraviolet oxidative dissolution -
ultraviolet absorption spectrophotometry
Total phosphorus: Potassium persulfate, ultraviolet oxidative dissolution -
Molybdenum blue spectrophotometric method
Number of
measurement
points
One point (standard), two points (option)
Number of
measurement
ranges
*3
One range (standard), 2 ranges (option)
Measurement
cycle
60 min
Installation
conditions
Ambient
temperature
2
C to 40
C
Ambient humidity Relative humidity 85
or less (no condensation)
Ambient
environment
Indoor installation type
Transient overvoltage of main power supply: Overvoltage Category II (EN61010-
1), Pollution Degree 2
Flat surface that is not subject to vibration and external shock
No dust, mist, or corrosive gas is present in the atmosphere
Under atmospheric pressure
No direct sunlight
Well-ventilated area
Measurement
conditions
Sample water
conditions
*4, *5, *6 , *7
Temperature
2
C to 40
C (no freezing)
Flow rate
0.5 L/min to 5 L/min (when OF-3 is used)
5 L/min to 20 L/min(when OF-30 is used)
Pure water/tap
water supply
conditions
*8
Amount of usage 43 L/month (depending on the measurement range)
Water quality
When the customer supplies pure water, use the water of class
A2 or above, as specified in JIS K0557: 1998, which does not
contain nitrogen and phosphorus compounds.
When a pure water unit is used, the tap water criteria defined by
the Water Supply Act must be met.
Supply method
Supply by the tank (standard) or continuous supply by the pure
water equipment (option)
Supply pressure 0.1 MPa to 0.5 MPa
Содержание TPNA-500