Maintenance
192
1.
Remove the waste liquid nozzle unit from the tank.
2.
Remove the waste liquid tank.
3.
In a location where this operation can be performed safely, remove the tightening
cap of the waste liquid tank, and then remove the float sensor and waste liquid
nozzle unit.
4.
Perform the waste liquid processing for the waste liquid tank.
5.
After the processing, return the tank to its original condition.
Connector for float sensor
Waste liquid tank
Waste liquid tank
Waste liquid nozzle unit
Tightening cap
Float sensor