Troubleshooting
221
Alarm 5
Lo Temp P-cell
The temperature
sensor for the cell
block is faulty or
disconnected
The temperature
sensor is poorly
secured
Check the temperature data.
Contact HORIBA Advanced Techno.
“ INT. ANALOG
INPUT” (page
127)
Alarm 5
Lack Span
Insufficient solution
remaining in the span
solution bottle
Replace the span solution bottle.
Check that the span solution bottle is
correctly placed in the reagent tray.
“ Replacing the
reagent bottle”
(page 193)
Alarm 5
Lack Reag. A
to F
The remaining amount of
reagent in the reagent
bottle is insufficient
Replace the reagent bottle. Check that
the reagent bottle is correctly set in the
reagent tray.
“ Replacing the
reagent bottle”
(page 193)
Alarm 5
P ZERO Cal.
Incompatible pure
water is used
The measurement cell
is dirty
The analyzer
components are faulty
Replace pure water and pure water
cartridge.
Replace the reagent bottle.
Contact HORIBA Advanced Techno.
-
Alarm 5
N ZERO Cal.
Incompatible pure
water is used
The measurement cell
is dirty
The analyzer
components are faulty
Replace pure water and pure water
cartridge.
Replace the reagent bottle.
Contact HORIBA Advanced Techno.
-
Alarm 5
P SPAN Cal.
Incompatible span
solution is used
The measurement cell
is dirty
The analyzer
components are faulty
Replace the span solution bottle.
Replace the reagent bottle.
Contact HORIBA Advanced Techno. -
Alarm 5
N SPAN Cal.
Incompatible span
solution is used
The measurement cell
is dirty
The analyzer
components are faulty
Replace the span solution bottle.
Replace the reagent bottle.
Contact HORIBA Advanced Techno. -
Alarm 5
UV ZERO Cal.
Incompatible zero
solution is used
The measurement cell
is dirty
The analyzer
components are faulty
Be sure to use distilled water for the
zero span solution.
Clean the measurement cell.
If the wiper has deteriorated, replace
it.
-
Alarm 5
VIS ZERO Cal.
Incompatible zero
solution is used
The measurement cell
is dirty
The analyzer
components are faulty
Be sure to use distilled water for the
zero span solution.
Clean the measurement cell.
If the wiper has deteriorated, replace
it.
-
Alarm 5
UV SPAN Cal.
Incompatible span
solution is used
The measurement cell
is dirty
The analyzer
components are faulty
Use the dedicated span solution. Be
sure to use distilled water for the
dilution water.
Clean the measurement cell.
If the wiper has deteriorated, replace
it.
-
Classification
Alarm
description
Possible cause
Corrective action
Reference