Function
72
AUTO WASH SET (Not supported)
Specifies the cycle and start time of the automatic reagent line cleaning.
Tip
For the automatic reagent line cleaning process, the reagent E line is cleaned.
Reagent E has a tendency to crystallize. Specifying periodic automatic cleaning will enable stable
measurement to be performed continuously.
Note
After the automatic reagent line cleaning, measurement stands by for one hour.
Be sure that different dates are specified for the start date of automatic reagent cleaning and
automatic hydrochloric acid cleaning. If the same date is specified, a higher priority is always
assigned to the automatic reagent line cleaning.
If the same date is specified for the start date of the automatic reagent line cleaning and the
automatic calibration, both the automatic reagent line cleaning and automatic calibration are
performed simultaneously.
Example) When 1 time is specified for all the “Pre-Wash”, “ZERO CAL.”, “SPAN CAL.”, and “Post-
Wash” settings, the operation is performed in the order of automatic reagent line cleaning, pre-
wash, zero calibration, span calibration, and post-wash.
Operation procedure
1.
Press [AUTO WASH SET] in the OPERATION SET 1/2 screen.
The AUTO WASH SET screen will be displayed.
2.
Specify the cycle and start time of the automatic reagent line cleaning.
Item
Setting range
Default value
Description
Function
ON
OFF
OFF
Selects whether or not to use the setting.
Cycle
1 to 31
30 days
Determines the automatic cleaning cycle.
Start Date
2009/01/01 00 to
2099/12/31 23
2009/01/01 00
Specifies the initial automatic cleaning date.
[MENU]
Main
MENU
SETTING
OPERATION SET1/2
[OPERATION SET]
[SETTING]
Refer to “ To display the [MENU] item screen” (page 62).
Содержание TPNA-500
Страница 1: ...Automatic Total Nitrogen Phosphorus Monitoring System TPNA 500 Instruction Manual CODE GZ0000371324D...
Страница 9: ...Type 1 2 4 3 5 2 type labels 6 For automatic hydrochloric acid cleaning specification...
Страница 12: ......
Страница 134: ...Function 118 MAINTENANCE STEP STEP ACTION SET Select various settings DECIDE START ACTION...
Страница 260: ......