Function
79
CONC. ALARM SET
Specifies the TP/TN concentration alarm value.
Operation procedure
1.
Press [CONC. ALARM SET] in the MEAS. LINE SET screen.
The CONC. ALARM SET screen will be displayed.
Item
Setting range
Default value
Description
Hi Limit L1 to L2
0.0000 mg/L to 2000 mg/L
0.0000 mg/L
Specifies the upper limit alarm value for the
TP/TN concentration.
H.Hi Limit L1 to L2
0.0000 mg/L to 2000 mg/L
0.0000 mg/L
Specifies the final upper limit alarm value for
the TP/TN concentration.
Note 1: L2 can be selected only for the two-point meter specification.
Note 2: When these parameters are set to 0.0000, no alarm will be activated.
Main
MENU
SETTING
MEAS. LINE SET
[MENU]
[MEAS. LINE SET]
[SETTING]