Function
101
CLOCK ADJUST
Specifies the clock settings. Select whether to enable (ON) or disable (OFF) the function that
allows the clock time to be changed using an arbitrary time value.
Tip
The time cannot be modified during operation.
When the time is modified by the external contact input, the input can be accepted from
1 min to
1 min of the modification time, and the time is modified accordingly.
Operation procedure
1.
Press [SYSTEM SET] in the SETTING screen.
The SYSTEM SET 1/2 screen will be displayed.
2.
Press [CLOCK ADJUST].
The CLOCK ADJUST screen will be displayed.
3.
Specify the system settings.
Item
Setting range
Default value
Description
CLOCK ADJUST
2001/01/01 00:00 to
2099/12/31 23:59
Current time
The time entered here is set as the current
time.
Modify function
ON
OFF
Select whether to enable (ON) or disable
(OFF) the time modification function at the
desired timing.
OFF
Modify input
Modify Date
Meas. Start Time
When “Modify function” is set ON, the time is
modified when the contact for time modification
is input.
Meas. Start Time
When “Modify function” is set ON, the time is
modified when the contact for measurement
start is input.
Line Select L1 to L2
When “Modify function” is set ON, the time is
modified when the contact for line selection is
input.
Modify date
0 min to 59 min
0 min
When “Modify function” is set ON, the clock
time is changed to the specified modified time
value entered at this point.
Note: L2 can be selected only for the two-point meter specification.
[MENU]
Main
Refer to “ To display the [MENU] item screen” (page 62).
[SETTING]
SETTING
MENU