Function
131
[DATA] - [LOG DATA]
Calls up the past measurement data.
MEASURE VALUE
Calls up the past measurement data.
The measurement value data are listed in the order they were updated.
Operation procedure
1.
Press [LOG DATA] in the DATA 1/2 screen.
2.
Press [MEASURE VALUE] in the LOG DATA screen.
The MEASURE VALUE screen will be displayed.
Pressing [
↑
] or [
↓
] scrolls the display. Or, press the button of the relevant date button
and select the measurement date you wish to view and press [DECIDE]. The data on
the specified date will be displayed.
To change the measurement element or line, press the button of the measurement
element/line and select the element/line you wish to view. Then, press [DECIDE].
To print, press [PRINT]. The list displayed on the screen will be printed out.
Item
Description
Reference
Displays the past measurement data in the order they
were updated.
Page 131
Displays the past hourly report data obtained every hour
in chronological order.
Page 133
[DATA]
Refer to “ To display the [DATA] item screen” (page 128).
DATA 1/2
Main
Item
Setting range
Measurement
element
TP
TN
-
Line
L1
L2
ZERO
SPAN
Date
2009/01/01 00 to 2099/12/31 23
Note: L2 can be selected only for the two-point meter specification.