Function
135
[DATA] - [GRAPH]
Displays the graph of the following data. The display setting of each graph can also be specified.
MEASURE VALUE
Displays the graph of the past measurement data. The display settings of the graph can be specified.
Operation procedure
1.
Press [GRAPH] in the DATA 1/2 screen.
2.
Press [MEASURE VALUE] in the GRAPH screen.
The MEASURE VALUE (graphical display) screen will be displayed.
Tip
Pressing [
←
]/[
→
] in the graphical display screen moves to the previous/next plot, allowing you to
check the data.
3.
To change the graphical display setting, press [SET] in the MEASURE VALUE
screen.
The GRAPH SET screen will be displayed.
4.
Specify the graphical display conditions on the GRAPH SET screen.
Item
Description
Reference
Displays the graph of the past measurement data. The
display settings of the graph can be specified.
Page 135
Displays the trend graph of the latest measurement data.
The display settings of the graph can be specified.
Page 137
HOUR REPORT HISTORY
Displays the graph of the past hourly report data. The
display settings of the graph can be specified.
Page 139
OPTICAL SIGNAL
Displays the graph of the past optical signal output per
measurement. The display settings of the graph can be
specified.
Page 141
TREND GRAPH (OPTICAL
SIGNAL)
Displays the real-time optical signal output. The display
settings of the graph can be specified.
Page 144
Item
Setting range
Setting
Measure Line
L1
L2
-
Specifies the line to be displayed in a graphical form.
Scale
0.01 to 2000
Specifies the graph display scale (concentration range).
Start Time
2009/01/01 00 to 2099/12/31 23
Specifies the display start date.
Period
1 day
7 days
30 days
-
Specifies the display range (period).
Vertical Top
25%
50%
75%
100%
Specifies the upper limit on the vertical axis of the graph
for the full scale of the measurement range.
Vertical Under
0%
25%
50%
75%
Specifies the lower limit on the vertical axis of the graph
for the full scale of the measurement range.
Meas. Element
TP
TN
-
Specifies TP/TN of the graph.
Note: L2 can be selected only for the two-point meter specification.
[DATA]
Refer to “ To display the [DATA] item screen” (page 128).
DATA 1/2
Main
Содержание TPNA-500
Страница 1: ...Automatic Total Nitrogen Phosphorus Monitoring System TPNA 500 Instruction Manual CODE GZ0000371324D...
Страница 9: ...Type 1 2 4 3 5 2 type labels 6 For automatic hydrochloric acid cleaning specification...
Страница 12: ......
Страница 134: ...Function 118 MAINTENANCE STEP STEP ACTION SET Select various settings DECIDE START ACTION...
Страница 260: ......