Connection
24
Contact input
This equipment has the following built-in contact input items.
Each input item is enabled when the relevant contact is set, and the following operation also
becomes available.
Note
The contact input selection is configured at the factory according to the customer's specifications.
Input item
Standard specification
Only available for 3-component specification
Item
Description
Input condition
Start signal type
For level setting
For pulse
setting
Meas. Start
Starts measurement.
The contact is ON (short-circuited)
for at least one second after the
OFF (release) status.
Level reading
Pulse reading
Cal. Start
Starts calibration.
Pulse reading
Pulse reading
Cleaning Start
Starts cleaning.
Pulse reading
Pulse reading
Modify Date
Accepts Modify Date.
Pulse reading
Pulse reading
Full Drain Tank
Accepts Full Drain
Tank.
The contact is ON (short-circuited)
for at least 10 seconds after the OFF
(release) status.
Level reading
Level reading
Samp. Lack L1
Accepts Samp. Lack
L1.
Level reading
Level reading
Line Select L1
Accepts Line Select
L1.
The signal is accepted when the
Meas. Start signal is input.
The contact is ON (short-
circuited) for at least one second
after the OFF (release) status.
Level reading
Level reading
Flow Mainte L1
Accepts Flow Mainte
L1.
The contact is ON (short-circuited)
for at least 10 seconds after the OFF
(release) status.
Level reading
Level reading
Flow Err. L1
Accepts Flow Err. L1.
Level reading
Level reading
Flow Power L1
Accepts Flow Power
L1.
Level reading
Level reading
Flow No Drain
L1
Accepts Flow No Drain
L1.
Level reading
Level reading
No allocation
-
-
-
-
Item
Description
Input condition
Start signal type
For level setting
For pulse
setting
UV/COD
Mainte L1
Accepts UV/COD
Mainte L1.
The contact is ON (short-circuited)
for at least 10 seconds after the OFF
(release) status.
Level reading
Level reading
UV/COD Err.
L1
Accepts UV/COD Err.
L1.
Level reading
Level reading
UV/COD
Power L1
Accepts UV/COD
Power L1.
Level reading
Level reading