Function
116
EXT. ANALOG OUTPUT ADJUST
Specifies the external analog output adjustment.
Operation procedure
1.
Press [EXT. ANALOG OUTPUT ADJUST] in the MAINTENANCE screen.
2.
The EXT. ANALOG OUTPUT ADJUST screen will be displayed.
3.
Specify the external analog output adjustment settings.
Note
After approximately 10 minutes have passed with no button operation occurring on the screen within
the
outlined area shown in the illustration below, output will revert to the setting specified
for ch.
Item (zero)
Setting range
Default value
Description
ch00
0001 to 0FFE
03D0
ch00 Adjusts the TP concentration analog zero output.
ch00
0001 to 0FFE
03D0
ch01 Adjusts the TN concentration analog zero output.
ch02
0001 to 0FFE
03D0
ch02 Adjusts the TP load analog zero output.
ch03
0001 to 0FFE
03D0
ch03 Adjusts the TN load analog zero output.
Item (span)
Setting range
Default value
Description
ch00
0001 to 0FFE
0DD4
ch00 Adjusts the TP concentration analog span output.
ch00
0001 to 0FFE
0DD4
ch01 Adjusts the TN concentration analog span output.
ch02
0001 to 0FFE
0DD4
ch02 Adjusts the TP load analog span output.
ch03
0001 to 0FFE
0DD4
ch03 Adjusts the TN load analog span output.
[MENU]
Main
Refer to “ To display the [MENU] item screen” (page 62).
[SETTING]
MAINTENANCE
MENU
[EXT.
MAINTENANCE
DA ADJ. ZERO
DA ADJ. SPAN
[SPAN]
[ZERO]
<Select
<Select
ANALOG
OUTPUT
ADJUST]
various
settings>
various
settings>
[ ]
[ ]