Function
74
OFF-LINE SET
Specifies the off-line measurement.
Tip
In the off-line measurement mode, a bottled sample is measured.
When the off-line mode is set to ON, the sample is not discharged from the sampling tube but
discharged from the drain tube of the overflow tank. Due to this, the bottled sample is not affected.
The amount of sample water to be consumed per measurement differs depending on the
measurement range (refer to the table below). Other operations are the same as those during normal
operation.
Operation procedure
1.
Press [OFF-LINE SET] in the OPERATION SET 2/2 screen.
The OFF-LINE SET screen will be displayed.
2.
Specify the off-line measurement.
Item
Setting range
Default value
Description
OFF-LINE MODE
ON
OFF
OFF
Specifies whether or not to perform the off-line
measurement.
TP measurement
range
(mg/L)
TN measurement range (mg/L)
2
5
10
20
50
100
200
500
1000
0.5
100
100
65
65
65
1
100
100
65
65
65
65
2
65
65
30
30
30
30
30
5
65
65
30
30
30
30
30
30
10
65
65
30
30
30
30
30
30
30
20
65
30
30
30
30
30
30
30
50
30
30
30
30
30
30
30
100
30
30
30
30
30
30
200
30
30
30
30
30
Unit: mL
[MENU]
Main MENU SETTING OPERATION SET 1/2
OPERATION SET 2/2
[OPERATION SET]
[SETTING]
Refer to “ To display the [MENU] item screen” (page 62).
[
→
]
OFF-LINE
MODE
OPERATION SET 2/2
OFF-LINE SET
[OFF-LINE
SET]
[ ]