Function
95
MAINTE OUTPUT SET
Specifies the maintenance output.
Operation procedure
1.
Press [DATA OUTPUT SET] in the OUTPUT SET screen.
The DATA OUTPUT screen will be displayed.
2.
Press [MAINTE OUTPUT SET].
The MAINTE OUTPUT screen will be displayed.
3.
Specify the maintenance output.
Item
Setting range
Default value
Description
Output Mode
Hold
Hold
Locks the TP/TN concentration analog output during
maintenance to the previous value.
Preset
Locks the TP/TN concentration analog output during
maintenance to the preset value.
Direct
Outputs the measurement value to the TP/TN
concentration analog output during maintenance.
Preset
0 mA to 22 mA
22 mA
Specifies the preset value for the TP/TN concentration
analog output during maintenance.
Main
MENU
SETTING
OUTPUT SET
[MENU]
[OUTPUT SET]
[SETTING]
Refer to “ To display the [MENU] item screen” (page 62).
[DATA
DATA OUTPUT
OUTPUT SET
[MAINTE
OUTPUT
SET]
OUTPUT
SET]
Preset
Output
Output Mode
Preset
MAINTE OUTPUT
<Input a value>
Mode
[ ]
[ ]