Connection
23
Contact output standard
The contact output standard is shown below.
Note
Do not apply any load exceeding the maximum rating. It could cause malfunction.
If the contact for the load is opened/closed, connect a spark killer, surge absorber (AC or DC
load), or diode (DC load), etc. in parallel with the load in order to prevent noise generation.
Signal type
Input/output circuit
Standard
Contact
output signal
Contact rating
250 V AC 3 A
30 V DC 3 A
(Note: However, only 30 V DC 3 A is available for the “Maintenance” contact.)
Contact output
No-voltage output contact
(When the allocated operation is activated, the contacts between terminals
are closed.)
Maintenance