![horiba TPNA-500 Скачать руководство пользователя страница 248](http://html1.mh-extra.com/html/horiba/tpna-500/tpna-500_instruction-manual_2162279248.webp)
Material
232
Hydrochloric acid consumption (page 46)
Sample water consumption (page 74)
Pure water supply cycle (guide) (page 204)
Cleaning cycle
1 hour
2 hours
3 hours
4 hours
6 hours
12
hours
24
hours
Monthly consumption (L)
18.6
9.3
6.2
4.7
3.1
1.6
0.8
4 L tank replacement
cycle (number of days)
7
13
20
27
40
80
160
TP measurement
range
(mg/L)
TN measurement range (mg/L)
2
5
10
20
50
100
200
500
1000
0.5
100
100
65
65
65
1
100
100
65
65
65
65
2
65
65
30
30
30
30
30
5
65
65
30
30
30
30
30
30
10
65
65
30
30
30
30
30
30
30
20
65
30
30
30
30
30
30
30
50
30
30
30
30
30
30
30
100
30
30
30
30
30
30
200
30
30
30
30
30
Unit: mL
TP measurement
range (mg/L)
TN measurement range (mg/L)
2
5
10
20
50
100
200
500
1000
0.5
21
21
19
19
16
-
-
-
-
1
21 21 19 19 16 18 -
-
-
2
19 19 18 18 15 16 14 -
-
5
19 19 18 18 15 16 14 12 -
10
16 16 15 15 13 14 12 11 11
20
-
18 16 16 14 15 13 11 11
50
-
-
14 14 12 13 11 10 10
100
-
-
- 12 11 11 10 9 9
200
-
-
-
-
11
11
10
9
9
Unit: Day
Содержание TPNA-500
Страница 1: ...Automatic Total Nitrogen Phosphorus Monitoring System TPNA 500 Instruction Manual CODE GZ0000371324D...
Страница 9: ...Type 1 2 4 3 5 2 type labels 6 For automatic hydrochloric acid cleaning specification...
Страница 12: ......
Страница 134: ...Function 118 MAINTENANCE STEP STEP ACTION SET Select various settings DECIDE START ACTION...
Страница 260: ......