Preparation for Operation
42
3.
Check that the pipings outside are complete, then close the V-1 valve and open the
V-3 valve for leading tap water to purge inside of the pipings.
Basic purge time ; about 5 minutes (Depending on the workmanship of the pipe
construction)
4.
Press
【
MENU
】
-
【
MAINTENANCE
】
-
【
SEPARATE ACTION
】
-
【→】
on the screen.
The SEPARATE ACTION 5/5 menu will be opened.
5.
Press
【
Pure water purge
】
to start purging of the pure water cartridge.
Basic purge time ; about 5 minutes
(Depending on the water quality of the tap water)
V-3 valve
V-1 valve
【
MENU
】
【
MAINTENANCE
】
Main screen
MENU
MAINTENANCE
【
SEPARATE ACTION
】
SEPARATE ACTION 5/5
【→】