Function
98
Measurement value output and setting
PAYLOAD SET
Specifies the load calculation.
Operation procedure
1.
Press [PAYLOAD SET] in the SETTING screen.
The PAYLOAD SET screen will be displayed.
2.
Specify the load calculation.
Tip
When “Date (Hourly)” is set to 59 min, hourly report is calculated based on the concentration
measured from 00 min to 59 min and the accumulated flow rate from 00 min to 59 min. Then, the
results are output at the point of 59 min.
When “Date (Daily)” is set to 23:00, the hourly report from 0:00 to 23:00 hours and data in the
hourly report is accumulated. Then, a daily report is output at 23:59.
Setting
Minus Data
OFF
ON
OFF
ON
OFF
ON
OFF
ON
Full-scale Over
OFF
OFF
ON
ON
OFF
OFF
ON
ON
Output Limit
100
100
100
100
110
110
110
110
Analog output
0% to 100
0% to 110
Display/printing
0 to F.S
Actual
measure-
ment
value to
F.S
0 to
actual
measure-
ment
value
Actual
measure-
ment
value
0 to F.S
Actual
measure-
ment
value to
F.S
0 to
actual
measure-
ment
value
Actual
measure-
ment
value
Item
Setting range
Default value
Description
Function
ON
OFF
OFF
Select whether or not to use the
setting.
Date (Hourly)
0 min to 59 min
59 min
Set the clock time (minute).
Date (Daily)
00:00 to 23:00
23:00
Set the clock time (hour).
Unit
kg•m
3
kg•m
3
Specify the unit.
g•m
3
g•L
Flow F.S. L1, L2
0.01 to 999999
1000 m
3
/h
Specify the full scale.
Note 1: L2 can be selected only for the two-point meter specification.
Note 2: If you have changed any of the “Date (Hourly)”, “Date (Daily)”, and “Unit” settings, the daily
report is output at that point and summary for load calculation is restarted from zero.
[MENU]
Main
Refer to “ To display the [MENU] item screen” (page 62).
[SETTING]
SETTING
MENU
Содержание TPNA-500
Страница 1: ...Automatic Total Nitrogen Phosphorus Monitoring System TPNA 500 Instruction Manual CODE GZ0000371324D...
Страница 9: ...Type 1 2 4 3 5 2 type labels 6 For automatic hydrochloric acid cleaning specification...
Страница 12: ......
Страница 134: ...Function 118 MAINTENANCE STEP STEP ACTION SET Select various settings DECIDE START ACTION...
Страница 260: ......