195
2.
As instructed on the screen, remove the reagent E bottle, and then replace it with
the reagent F bottle. Loosen the tightening cap of the reagent E bottle and remove
it together with the nozzle unit.
Note
As the reagent E bottle will be used in a later step, store it temporarily.
3.
Remove the reagent F bottle in the same manner as step 2. above, and then attach
it to line E.
4.
Once the procedure is completed, press [YES] on the screen.
Nozzle unit
Tightening cap
Reagent E bottle
Tube: Orange
Label: Orange
Reagent E bottle
Reagent F bottle
Tube: Orange
Label: Red
Label: Orange
Line E
(Replace the reagent E bottle with the
reagent F bottle)