Function
149
About the data to be stored
Any data stored in the TPNA-500 can be transferred to a USB memory device so that you can
check it on a PC, etc. In order to check the data at a later time, it can also be printed out on the
printer.
Available data to be transferred to a USB
Measurement value (MES.CSV)
For TP/TN
“Optical signal (PUL5.CSV)” “ABS.3” is enabled only for turbidity compensation function
specification.
Description
Maximum number of units to be stored
Reference
Measurement value
(MES.CSV)
Stores 26280 latest values for each element.
Page 149
Hourly report (HRS.CSV)
Stores 8760 latest values for each line and element.
Page 150
Daily report (DAY.CSV)
Stores 365 latest values for each line and element.
Page 151
Alarm history (ALM.CSV)
Stores 500 latest events.
Page 151
Calibration history (CAL.CSV)
Stores 100 latest events.
Page 152
Pressure signal (PUL1.CSV)
Stores 75 latest pressure signal values.
Page 153
Optical signal (PUL2.CSV)
Stores 75 latest REF signal values for TN.
Optical signal (PUL3.CSV)
Stores 75 latest SMP signal (220 nm) values for TN.
Optical signal (PUL4.CSV)
Stores 75 latest SMP signal values for TP.
Optical signal (PUL5.CSV)
Stores 75 latest SMP signal (275 nm) values for TN.
Note: This data are not available in this specification.
1-minute value (MIN.CSV)
Stores 14400 latest UV meter values.
Note: Only for the UV meter
Page 154
Operation history (OPE.CSV)
Stores the history when an operation is performed by the user or the
equipment mode is switched.
Stores 100 latest events.
Page 154
Setting values (USER.CSV)
Stores the latest setting values for each setting item.
-
Header
Description
YY/MM/DD/HH:MM:SS
Measurement time
LINE
Measurement line
ELEMENT
Measurement element
CONC
Concentration
REAL CONC
Concentration (before revision)
REVISON A
Revision parameter A
REVISON B
Revision parameter B
CONC ALARM
Concentration alarm
ABS.1
Absorbance (220 nm)
ABS.2
Inside temperature of the equipment (
C)
ABS.3
Absorbance (275 nm)
ABS.1 ALARM
Absorbance alarm
ABS.2 ALARM
---
ABS.3 ALARM
---