Material
243
Operation procedure
1.
Press [TURB. COEFFICIENT SET] in the MEAS. LINE SET2/2 screen.
The TURB. COEFFICIENT SET screen will be displayed.
2.
Specify the coefficient for the turbidity compensation.
【
DATA
】
-
【
SAVE TO USB
】
When the turbidity compensation is effective, the maximum number of the saving data will be
different from the number of "turbidity compensation OFF".
Refer to the following table.
Available data to be transferred to a USB
Main
MENU1/2
SETTING
MEAS. LINE SET2/2
[MENU]
[MEAS. LINE SET]
[SETTING]
Refer to “ To display the [MENU] item screen” .
TURB.COEFFICIENT SET
MEAS. LINE SET2/2
Coefficient
【
TURB.
Coefficient L1
【 】
《
Input a value
》
COEFFICIENT
SET
】
L1
Description
Maximum number of units to be stored
Measurement value (MES.CSV) Stores 26280 latest values for each element.
Hourly report (HRS.CSV)
Stores 8760 latest values for each line and element.
Daily report (DAY.CSV)
Stores 365 latest values for each line and element.
Alarm history (ALM.CSV)
Stores 500 latest events.
Calibration history (CAL.CSV)
Stores 100 latest events.
Pressure signal (PUL1.CSV)
Stores 60 latest pressure signal values.
Optical signal (PUL2.CSV)
Stores 60 latest REF signal values for TN.
Optical signal (PUL3.CSV)
Stores 60 latest SMP signal (220 nm) values for TN.
Optical signal (PUL4.CSV)
Stores 60 latest SMP signal values for TP.
Optical signal (PUL5.CSV)
Stores 60 latest SMP signal (275 nm) values for TN.
1-minute value (MIN.CSV)
Stores 14400 latest UV meter values.
Note: Only for the UV meter
Operation history (OPE.CSV)
Stores the history when an operation is performed by the user or the equipment mode
is switched.
Stores 100 latest events.
Setting values (USER.CSV)
Stores the latest setting values for each setting item.