Material
242
TURB. COEFFICIENT SET (Option)
This section is the supplementary explanation for the turbidity compensation of the TPNA-500
when TN is measured.
MEAS. LINE SET
One line is added to the original table for the turbidity compensation.
TURB. COEFFICIENT SET
Select the turbidity compensation coefficient setting.
Note
The absorbance at 220 nm for TN measurement is influenced by turbidity, the absorbance at 270 nm
with a compensation factor is subtracted from the absorbance at 220 nm to obtain the turbidity-
compensated total nitrogen.
The turbidity compensation setting is not necessary to be changed under the condition that the good
correlation with the manual analysis is obtained.
Item
Description
LINE CHANGE SET
Specifies the allocation of the measurement line.
CONV. COEF. SET
Specifies the conversion coefficient.
CONC. ALARM SET
Specifies the TP/TN concentration alarm value.
LOAD ALARM SET
Specifies the TP/TN load alarm value.
FLOW ALARM SET
Specifies the flow rate alarm value.
TURB. COEFFICIENT SET
Specifies the turbidity coefficient.
Item
Setting range
Default value
Description
Coefficient L1
~
L2
0.000
~
3.000
2.000
Turbidity compensation coefficient setting.
Note : L2 can be selected only for the two-point meter specification.