Function
64
OPERATION MODE SET
Specifies the operation mode.
Operation procedure
1.
Press [OPERATION MODE SET] in the OPERATION SET 1/2 screen.
The OPERATION MODE screen will be displayed.
2.
Specifies the operation mode.
Item
Setting range
Default
value
Description
Control Mode
Local
Local
No external control
Remote (contact)
Controls measurement by the contact input.
Remote (communication)
Controls measurement by the communication commands.
Start Signal
Pulse
Pulse
Starts measurement by the pulse input.
Level
Performs measurement according to the measurement mode
while the signal is input.
Measure Mode
Schedule
Repeat
Performs measurement according to the specified schedule.
Repeat
Performs measurement repeatedly.
Meas. Element
TP
ON
Measures the TP.
TN
Measures the TN.
[MENU]
MEAS.
MENU
SETTING
OPERATION SET1/2
[OPERATION SET]
[SETTING]
Refer to “ To display the [MENU] item screen” (page 62).
Control Mode
Start Signal
Measure Mode
OPERATION SET 1/2
OPERATION MODE
Control Mode
Start Signal
Measure Mode
[OPERATION
[ ]
Meas.
ANALYZE TYPE SET
MODE SET]
[ ]
[ ]
[ ]
Element
[ ]
Meas. Element