Troubleshooting
214
Troubleshooting
When measurement cannot be started
If measurements cannot be started, the following are the possible causes.
Take corrective action as instructed below.
Symptom
Possible cause
Corrective action
Reference
Power is not turned
ON
The power is not being
supplied
Turn both the main power switch and
power switch ON.
Check that the power cable is correctly
connected.
“ Main power
switch (Ground
fault
interrupter)”
(page 2)
An alarm is
activated
A previously activated alarm
has not been canceled
Take corrective action as instructed in the
reference section.
“ Cause and
corrective
action for
alarms” (page
218)
Measurements
cannot be started
The operation settings are
incorrect
Perform the appropriate operation settings.