Troubleshooting
218
Cause and corrective action for alarms
The following is a list of alarms and corrective actions.
When corrective action is required, make sure you understand the instructions and then
perform the operation.
Classification
Alarm
description
Possible cause
Corrective action
Reference
Alarm 1
Factory Setting
The default setting value
has been initialized or
data for default setting
values stored on the F-
ROM has been corrupted
Turn OFF the power, and then turn it
ON again. If the problem cannot be
resolved, contact HORIBA Advanced
Techno.
-
Alarm 1
User Setting
The user setting value
has been initialized or
data for user setting
values stored on the F-
ROM has been corrupted
Turn OFF the power, and then turn it
ON again. If the problem cannot be
resolved, contact HORIBA Advanced
Techno.
-
Alarm 1
Measure
History
History data stored in CF
memory has been
corrupted
Turn OFF the power, and then turn it
ON again. If the problem cannot be
resolved, contact HORIBA Advanced
Techno.
-
Alarm 1
Hour Report
History data stored in CF
memory has been
corrupted
Turn OFF the power, and then turn it
ON again. If the problem cannot be
resolved, contact HORIBA Advanced
Techno.
-
Alarm 1
Day Report
History data stored in CF
memory has been
corrupted
Turn OFF the power, and then turn it
ON again. If the problem cannot be
resolved, contact HORIBA Advanced
Techno.
-
Alarm 1
Alarm History
History data stored in CF
memory has been
corrupted
Turn OFF the power, and then turn it
ON again. If the problem cannot be
resolved, contact HORIBA Advanced
Techno.
-
Alarm 1
Calib. History
History data stored in CF
memory has been
corrupted
Turn OFF the power, and then turn it
ON again. If the problem cannot be
resolved, contact HORIBA Advanced
Techno.
-
Alarm 1
Optical signal
History data stored in CF
memory has been
corrupted
Turn OFF the power, and then turn it
ON again. If the problem cannot be
resolved, contact HORIBA Advanced
Techno.
-
Alarm 1
OperateData
History data stored in CF
memory has been
corrupted
Turn OFF the power, and then turn it
ON again. If the problem cannot be
resolved, contact HORIBA Advanced
Techno.
-
Alarm 1
1-min.Data
History data stored in CF
memory has been
corrupted
Turn OFF the power, and then turn it
ON again. If the problem cannot be
resolved, contact HORIBA Advanced
Techno.
-
Alarm 1
EEPROM
Read
The EEPROM storage
data is damaged
Turn OFF the power, and then turn it
ON again. If the problem cannot be
resolved, contact HORIBA Advanced
Techno.
-
Alarm 1
Clock Err.
Clock backup battery
LOW
Meter clock failure
Contact HORIBA Advanced Techno.
-